SCPI Command
CALCulate<Ch>[:SELected]:DATA:FDATa?
Description
Reads out the formatted data array.
The formatted data array is the data, whose processing is completed including the formatting as the last step. Such data represent the data trace values as they are shown on the screen.
The array size is 2N, where N is the number of measurement points.
For the n–th point, where n from 1 to N:
<numeric 2n–1> Value 1 depends on the trace format (see table below);
<numeric 2n> Value 2 depends on the trace format (see table below).Reads out or writes the formatted data array.
Trace Format
|
Value 1
|
Value 2
|
Log Mag
|
Logarithmic magnitude, dB
|
0
|
SWR
|
Voltage standing wave ratio
|
0
|
Phase
|
Phase, deg
|
0
|
Expand Phase
|
Expanded phase, deg
|
0
|
Group Delay
|
Group delay, sec
|
0
|
Lin Mag
|
Linear magnitude
|
0
|
Real
|
Real part
|
0
|
Imag
|
Imaginary part
|
0
|
Smith (Log/Phase)
|
Logarithmic magnitude, dB
|
Phase, deg
|
Smith (Lin/Phase)
|
Linear magnitude
|
Phase, deg
|
Smith (Real/Imag)
|
Real part
|
Imaginary part
|
Smith (R + jX)
|
Impedance (real part), Ohm
|
Impedance (imaginary part), Ohm
|
Smith (G + jB)
|
Admittance (real part), S
|
Admittance (imaginary part), S
|
Polar (Log/Phase)
|
Logarithmic magnitude, dB
|
Phase, deg
|
Polar (Lin/Phase)
|
Linear magnitude
|
Phase, deg
|
Polar (Real/Imag)
|
Real part
|
Imaginary part
|
|
query only
|
Target
The active trace of channel <Ch>,
<Ch>={[1]|2|3|4}
<Ch>={[1] … |16} (in N-port mode only)
Query Response
<numeric 1>, <numeric 2>, …<numeric 2N>
Related Commands
CALC:FORM
FORM:DATA
Equivalent Softkeys
None
Equivalent COM Command
SCPI.CALCulate(Ch).SELected.DATA.FDATa
Syntax
Data = app.SCPI.CALCulate(Ch).SELected.DATA.FDATa
Type
Variant (array of Double) (read)
Back to CALCulate
Back to Command Finder
Back to Command Finder (From User Interface)