Please enable JavaScript to view this site.

 Network Analyzers using PXI software

Depending on the sweep settings the calibration commands may have a long execution time, as they start the sweep and wait for it to complete. These commands are:

SENS:CORR:COLL:XXXX

SENS:CORR:OFFS:COLL:XXXX

SENS:CORR:REC:COLL:XXXX

SENS:CORR:COLL:ECAL:XXXX

SENS:CORR:COLL:ECAL:ORI:EXEC

The user program can stop execution until the end of these commands using any query, the *OPC? for example.

 

Rev.:  22.3