Please enable JavaScript to view this site.

 Network Analyzers using PXI software

SCPI Command

SOURce<Ch>:POWer:CENTer <power>

SOURce<Ch>:POWer:CENTer?

Description

Sets or reads out the center value of the power sweep type.

command/query

Target

Channel <Ch>,

<Ch>={[1]|2|...16}

Parameter

<power> the power level within the power limits of the analyzer.

Unit

dBm (decibels above 1 milliwatt)

Resolution

0.05 dBm

Out of Range

Sets the value of the limit, which is closer to the specified value.

Query Response

<numeric>

Preset Value

Depends on the Analyzer

Equivalent Softkeys

Stimulus > Center

Back to SOURce

Back to Command Finder

Back to Command Finder (From User Interface)

 

Rev.:  22.3