Please enable JavaScript to view this site.

 Network Analyzers using TRVNA software

SCPI Command

MMEMory:STORe:SNP[:DATA] <string>

Description

Saves the measured S-parameters of the active channel into a Touchstone file. The file type (one-port or two-port) is set by the following commands: MMEM:STOR:SNP:TYPE:S1P, MMEM:STOR:SNP:TYPE:S2P. One–port type file saves one reflection parameter: S11.  Two–port type file saves two parameters: S11, S21.

Note: If the full path of the file is not specified, the \FixtureSim subdirectory of the application directory will be searched. The file has *.SNP extension by default.

no query

Target

Active channel, set by the DISP:WIND:ACT command.

Parameter

<string> file name (quoted string)

Equivalent Softkeys

System > Save > Touchstone File > Save


Equivalent COM Command

SCPI.MMEMory.STORe.SNP.DATA

Syntax

app.SCPI.MMEMory.STORe.SNP.DATA = File

Type

String (write only)


 

Back to MMEMory

Back to Command Finder

Back to Command Finder (From User Interface)

 

Rev.:  24.1