Please enable JavaScript to view this site.

 Network Analyzers using TRVNA software

SCPI Command

DISPlay:WINDow<Ch>:TRACe<Tr>:Y[:SCALe]:PDIVision <response>

DISPlay:WINDow<Ch>:TRACe<Tr>:Y[:SCALe]:PDIVision?

Description

Sets or reads out the trace scale. Sets the scale per division when the data format is in the rectangular format. Sets the full-scale value when the data format is in the Smith chart format or the polar format.

command/query

Target

Trace <Tr> of channel <Ch>,

<Tr>={[1]|2|...8}

<Ch>={[1]|2|...9}

Parameter

<response> the scale value

Out of Range

Sets the value of the limit, which is closer to the specified value.

Unit

dB |° |s

Query Response

<numeric>

Preset Value

Varies depending on the format.

Logarithmic Magnitude

10 dB/Div

Phase

40 °/Div

Expand Phase

100 °/Div

Group Delay

10e–9 s/Div

Smith Chart, Polar, SWR

1 /Div

Linear Magnitude

0.1 /Div

Real part, Imaginary part

0.2 /Div

Equivalent Softkeys

Scale > Scale


Equivalent COM Command

SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.PDIVision

Syntax

Value = app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.PDIVision

app.SCPI.DISPlay.WINDow(Ch).TRACe(Tr).Y.SCALe.PDIVision = 20

Type

Double (read/write)


 

Back to DISPlay