Please enable JavaScript to view this site.

 Network Analyzers using TRVNA software

SCPI Command

CALCulate<Ch>:PARameter:COUNt <numeric>

CALCulate<Ch>:PARameter:COUNt?

Description

Sets or reads out the number of traces in the channel.

command/query

Target

The channel <Ch>,

<Ch>={[1]|2|...9}

Parameter

<numeric> The number of the traces in the channel from 1 to 16

Out of Range

Sets the value of the limit, which is closer to the specified value.

Query Response

<numeric>

Preset Value

1

Equivalent Softkeys

None


Equivalent COM Command

SCPI.CALCulate(Ch).PARameter.COUNt

Syntax

TraceNum = app.SCPI.CALCulate(Ch).PARameter.COUNt

app.SCPI.CALCulate(Ch).PARameter.COUNt = 2

Type

Long (read/write)


 

Back to CALCulate

Back to Command Finder

Back to Command Finder (From User Interface)

 