Please enable JavaScript to view this site.

 Network Analyzers using S4VNA software

SCPI Command

SOURce<Ch>:POWer:PORT<Pt>:CORRection:COLLect[:ACQuire]

Description

Measures the power calibration data for the port <Pt> using the power meter controlled via USB or USB/GPIB. Calculates calibration coefficients on completion of the measurement and turns ON the power correction for the port.

query only

Target

Port <Pt> of channel <Ch>,

<Ch>={[1]|2|...16}

<Pt>={[1]|2|3|4}

Related Commands

SOUR:POW:PORT:CORR:COLL:TABL:LOSS

SOUR:POW:PORT:CORR:COLL:TABL:LOSS:DATA

Equivalent Softkeys

Calibration > Power Calibration > Take Cal Sweep


Equivalent COM Command

SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.COLLect.ACQuire

Syntax

app.SCPI.SOURce(Ch).POWer.PORT(Pt).CORRection.COLLect. ACQuire

Type

Method


 

Back to SOURce

Back to Command Finder

Back to Command Finder (From User Interface)

 

Rev.:  22.4