Please enable JavaScript to view this site.

 Network Analyzers using S4VNA software

SCPI Command

SOURce<Ch>:AUXiliary:FREQuency:STARt <numeric>

SOURce<Ch>:AUXiliary:FREQuency:STARt?

Description

Sets or reads out the start of the frequency range of the auxiliary RF source. When set the multiplier and offset values are automatically corrected.

command/query

Target

Channel <Ch>,

<Ch>={[1]|2|...16}

Parameter

<numeric> start of the frequency range of the auxiliary RF source within the analyzer frequency range

Unit

Hz (Hertz)

Out of Range

Sets the value of the limit, which is closer to the specified value.

Query Response

<numeric>

Preset Value

1 GHz

Related Commands

SOUR:AUX:FREQ:STOP

Equivalent Softkeys

Stimulus > Auxiliary Source > Start


Equivalent COM Command

None


 

Back to SOURce

Back to Command Finder

Back to Command Finder (From User Interface)

 

Rev.:  22.4