Please enable JavaScript to view this site.

 Network Analyzers using S4VNA software

SCPI Command

*OPC?

Description

Reads out the "1" at the completion of all pending operations. The query blocks the execution of the user program until execution of all previous instructions.

The query *OPC? can be used for waiting for the end of a sweep initiated by the command TRIG:SING.

query only

Target

Analyzer

Query Response

1

Related Commands

TRIG:SING

Equivalent Softkeys

None


Equivalent COM Command

SCPI.IEEE4882.OPC

Syntax

Value = app.SCPI.IEEE4882.OPC

app.SCPI.IEEE4882.OPC = Dummy

Type

Long (read/write)


 

Back to IEEE488.2 Common Commands

Back to Command Finder

Back to Command Finder (From User Interface)