SCPI Command
TRIGger[:SEQuence][:IMMediate]
Generates a trigger signal and initiates a sweep under the following conditions: 1. Trigger source is set to the BUS (set by the command TRIG:SOUR BUS), otherwise an error occurs and the command is ignored. 2. Analyzer must be in the trigger waiting state, otherwise (the analyzer is in the measurement state or in the hold state) an error occurs, and the command is ignored. The command is completed immediately after the generation of the trigger signal (does not wait the end of a sweep). |
---|
no query |
TRIG:SOUR BUS
None
Equivalent COM Command
SCPI.TRIGger.SEQuence.IMMediate
app.SCPI.TRIGger.SEQuence.IMMediate
Method
Back to TRIGger
Back to Command Finder