SCPI Command
*TRG
Generates a trigger signal and initiates a sweep under the following conditions. 1. Trigger source is set to the BUS (set by the command TRIG:SOUR BUS), otherwise an error occurs and the command is ignored. 2. Analyzer must be in the trigger waiting state, otherwise (the analyzer is in the measurement state or hold state) an error occurs, and the command is ignored. The command is completed immediately after the generation of the trigger signal. |
---|
no query |
Analyzer
None
Equivalent COM Command
SCPI.IEEE4882.TRG
app.SCPI.IEEE4882.TRG
Method
Back to IEEE488.2 Common Commands
Back to Command Finder