Please enable JavaScript to view this site.

 Network Analyzers using S2VNA software

SCPI Command

*TRG

Description

Generates a trigger signal and initiates a sweep under the following conditions.

1. Trigger source is set to the BUS (set by the command TRIG:SOUR BUS), otherwise an error occurs and the command is ignored.

2. Analyzer must be in the trigger waiting state, otherwise (the analyzer is in the measurement state or hold state) an error occurs, and the command is ignored.

The command is completed immediately after the generation of the trigger signal.

no query

Target

Analyzer

Related Commands

TRIG:SOUR

INIT

INIT:CONT

Equivalent Softkeys

None


Equivalent COM Command

SCPI.IEEE4882.TRG

Syntax

app.SCPI.IEEE4882.TRG

Type

Method


 

Back to IEEE488.2 Common Commands

Back to Command Finder

Rev.:  24.2