Please enable JavaScript to view this site.

 Network Analyzers using S2VNA software

SCPI Command

SENSe:SWEep:PULSe:TRIGger:READy {OFF|ON|0|1}

SENSe:SWEep:PULSe:TRIGger:READy?

Description

Enables "Ready for Pulse" signal. The Analyzer outputs this signal to the "Pulse 1 Out" connector when the following conditions are met:

The synchronous pulse mode (Point-in-pulse or Pulse Profile modes) is active.

The pulse trigger source is set to external.

The Analyzer is ready for the next trigger pulse.

Note: S5180B model only.

command/query

Target

Analyzer

Parameter

{ON|1}

Enables "Ready for Pulse" signal to be routed to "Pulse 1 Out" connector.

{OFF|0}

Disables "Ready for Pulse" signal.

Query Response

{0|1}

Preset Value

0

Equivalent Softkeys

Stimulus > Pulse Meas > Ready for Pulse


Equivalent COM Command

None


 

Back to Pulse Measurement S5180B

Back to Command Finder (From User Interface)

 