Please enable JavaScript to view this site.

 Network Analyzers using S2VNA software

SCPI Command

SENSe<Ch>:OFFSet:RECeiver[:FREQuency]:OFFSet <frequency>

SENSe<Ch>:OFFSet:RECeiver[:FREQuency]:OFFSet?

Description

Sets or reads out the basic frequency range offset to get the receiver frequency when the frequency offset feature is ON and offset type is "SRCRcv".

command/query

Target

Channel <Ch>,

<Ch>={[1]|2|...16}

Parameter

<frequency> offset from –1e12 to 1e12

Unit

Hz (Hertz)

Out of Range

Sets the value of the limit, which is closer to the specified value.

Query Response

<numeric>

Preset Value

0

Related Commands

SENS:OFFS

SENS:OFFS:TYPE

Equivalent Softkeys

Stimulus > Frequency Offset > Receivers > Offset


Equivalent COM Command

SCPI.SENSe(Ch).OFFSet.RECeiver.FREQuency.OFFSet

Syntax

Value = app.SCPI.SENSe(Ch).OFFset.RECeiver.FREQuency.OFFSet

app.SCPI.SENSe(Ch).OFFset.RECeiver.FREQuency.OFFSet = 1e9

Type

Double (read/write)


 

Back to SENSe

Back to Command Finder

 

Rev.:  24.2