SCPI Command
SENSe<Ch>:CORRection:RECeiver<Pt>:COLLect:ACQuire <srcport>
Executes receiver calibration of both the test receiver and the reference receiver of the specified port <Pt>.The test receiver calibration uses port number <srcport> as the source port. The reference receiver calibration uses its own port <Pt> as the source port. Note: The command starts the measurement of the calibration data immediately regardless the trigger settings. The command blocks the execution of the subsequent commands until the completion of the measurement. |
---|
no query |
Port <Pt> of channel <Ch>,
<Ch>={[1]|2|...16}
<Pt>={[1]|2}
<srcport> |
The number of the source port from 1 to 2 (or 4) |
Calibration > Receiver Calibration > Calibrate Both
Equivalent COM Command
SCPI.SENSe(Ch).CORRection.RECeiver(Pt).COLLect.ACQuire
app.SCPI.SENSe(Ch).CORRection.RECeiver(Pt).COLLect.ACQuire = Src
Long (write only)