Please enable JavaScript to view this site.

 Network Analyzers using S2VNA software

SCPI Command

SENSe<Ch>:CORRection:EXTension:AUTO:STOP <frequency>

SENSe<Ch>:CORRection:EXTension:AUTO:STOP?

Description

Sets or reads out the stop value of the user span of the auto port extension function.

command/query

Target

Channel <Ch>,

<Ch>={[1]|2|...16}

Parameter

<frequency> the user span stop

Unit

Hz (Hertz)

Out of Range

Sets the value of the limit, which is closer to the specified value.

Query Response

<numeric>

Preset Value

The Analyzer's lowest frequency.

Related Commands

SENS:CORR:EXT:AUTO:CONF

Equivalent Softkeys

Calibration > Port Extension > Auto Port Extension > User Span Stop


Equivalent COM Command

None


 

Back to SENSe

Back to Command Finder

 

Rev.:  24.2