Please enable JavaScript to view this site.

 Network Analyzers using S2VNA software

For a one/two-port calibration, connect any ports of the ACM to the ports to be calibrated on the Analyzer.

AutoCal Module

To open automatic calibration submenu, use the following softkeys:

Calibration > AutoCal

1-port Autocal (2port)

To perform full one-port calibration, use the 1-Port AutoCal softkey.

Then, select the port number and click the softkey to start the calibration.

Wait until calibration is complete.

The instrument status bar will indicate Calibration in progress... when the measurement is in progress.

 

OnePath 2-port AutoCal

To perform one path two-port calibration, use the One Path 2-Port Auto softkey.

Then, select the direction of the calibration using the Port n->m softkey. The label on the softkey indicates the measured parameters.

Click Port n->m softkey to start the calibration.

Wait until calibration is complete.

The instrument status bar will indicate Calibration in progress... when the measurement is in progress.

 

2-Port AutoCal

To perform full two-port calibration, click the 2-Port AutoCal softkey.

Wait until calibration is complete.

The instrument status bar will indicate Calibration in progress... when the measurement is in progress.

 

 

 

 

 

Rev.:  24.2