Please enable JavaScript to view this site.

 Network Analyzers using PXI software

The definitions of the calibration standards included in one calibration kit are listed in the table as shown below.

Calibration standard definition table 1

Calibration standard definition table 2

Calibration standard definition table for the standards defined by the model

The Analyzer provides two methods of defining a calibration standard:

Calibration standard model

Table of S-parameters

The calibration standards defined by S-parameters are called Data-Based standards. For the Data-Based standards editing, see Data-Based Calibration Standards.

Each calibration standard is characterized by lower and upper values of the operating frequency. In the process of calibration, the measurements of the calibration standards outside the specified frequency range are ignored.

Cal Kit 8850Q

Define STDs

To open the table of calibration standard definitions, use the following softkeys:

Calibration > Cal Kit > Define STDs

Standard Adding to Calibration Kit

Add STD softkey

To add a calibration standard to the table of calibration standard definition (See figure above), use the following softkey:

Calibration > Cal Kit > Define STDs > Add STD

 

 

 

Standard Deleting from Calibration Kit

Delete STD

To delete a calibration standard from the table of calibration standard definition (See figure above), use the following softkey:

Calibration > Cal Kit > Define STDs > Delete STD

 

 

 

Calibration Standard Editing

Moving in the table of calibration standard definitions (See above figure) using navigation keys. Enter the parameter values for a calibration kit by using the navigation keys in the table of calibration standard definitions:

N

The calibration standard number is specified in the calibration kit data sheet (just for information).

Type

Select the standard type:

Open

Short

Load

Thru/Line

Unknown Thru

Sliding Load

Data-Based

Label

Standard labels specified on the calibration menu softkeys.

F min

Minimum operating frequency of the coaxial standard.

Lower cutoff frequency of the waveguide standard.

F max

Maximum operating frequency of the coaxial standard.

Upper cutoff frequency of the waveguide standard.

Delay

Offset delay value in one direction (s). Can be switched to physical length (m). The parameter is used only for the calibration standard model.

Z0

Offset characteristic impedance value (Ω).

For waveguide must be set to 1 Ω.

Loss

Offset loss value (Ω/s). The parameter is used only for the definition of the standard with the help of the calibration standard model.

Media

Coaxial or Waveguide

H/W

Waveguide height to width ratio.

Terminal Impedance

Lumped load impedance value (Ω). The parameter is used only for the definition of the standard with the help of the calibration standard model.

C0 10–15 F

For an OPEN standard, C0 coefficient in the polynomial formula of the fringe capacitance:

C1 10–27 F/Hz

For an OPEN standard, C1 coefficient in the polynomial formula of the fringe capacitance.

C2 10–36 F/Hz2

For an OPEN standard, C2 coefficient in the polynomial formula of the fringe capacitance.

C2 10–45 F/Hz3

For an OPEN standard, C3 coefficient in the polynomial formula of the fringe capacitance.

L0 10–12 H

For a SHORT standard, L0 coefficient in the polynomial formula of the residual inductance:

L1 10–24 H/Hz

For a SHORT standard, L1 coefficient in the polynomial formula of the residual inductance.

L2 10–33 H/Hz2

For a SHORT standard, L2 coefficient in the polynomial formula of the residual inductance.

L2 10–42 H/Hz3

For a SHORT standard, L3 coefficient in the polynomial formula of the residual inductance.

 

 

 

Calibration Standard Copy/Paste Function

Copy STD softkey

Copy All STDs softkey

To save a calibration standard into clipboard, highlight the required line in the calibration standard definition table, and click the following softkey:

Copy STD or Copy All STDs

Paste

To paste the standard(s) from the clipboard, click the following softkey:

Paste

Offset Delay Measurement Units Switching

Offset Unit Seconds

Seconds dot softkey

Meters softkey

To switch the offset delay measurement units in the calibration standard definition table (See figure above), click the following softkey:

Offset Unit > Seconds | Meters

Offset Permittivity 1.00649

 

To enter the offset permittivity, click the following softkey:

Offset Permittivity

The offset permittivity is used only for the delay to length conversion. Default value equals the permittivity of air.

Management of Sequence in Standard Table

STD Up

STD Down

To change the sequence in the table, use the following softkeys:

STD Up or STD Down

 

Rev.:  22.3