 Network Analyzers using PXI software

This section describes how to edit the calibration kit description, to add and delete a calibration kit.

The Analyzer provides a table for 64 calibration kits. The first part of the table contains the predefined kits. The second part of the table is for calibration kit added by the user.

A calibration kit redefining can be required for the following purposes:

To change the port assignment of a standard to ensure connector type (male, female) matching.

To add a user-defined standard into the kit, e.g. a non-zero-length thru.

To precise the standard parameters to improve the calibration accuracy.

A new user-defined calibration kit adding can be performed when a required kit is not included in the list of the predefined kits.

The deleting function is available for user-defined calibration kits only.

The restore function is available for predefined calibration kits only.

Any changes made to the calibration kits are automatically saved into the nonvolatile memory of the Analyzer. Clicking the Save button is not required in order to save.


note

Changes to a predefined calibration kit can be canceled at any time and the initial state will be restored.


 

Rev.:  22.3