Please enable JavaScript to view this site.

 CMT VNA Programming Manual

SCPI Command

TRIGger[:SEQuence]:POINt {OFF|ON|0|1}

TRIGger[:SEQuence]:POINt?

Description

Turns the point trigger feature ON/OFF.

When the point trigger is turned ON, the external trigger response is the single point. When the point trigger feature is turned OFF, the external trigger response is the entire sweep.

command/query

Parameter

Specifies the point trigger function state:

{ON|1}

ON

{OFF|0}

OFF

Query Response

{0|1}

Preset Value

0

Related Commands

TRIG:SOUR EXT

Back to TRIGger

Back to Command Finder

 

Rev.:  26.00