Please enable JavaScript to view this site.

 Vector Network Analyzers Using CMT VNA Software

One of four trigger sources can be selected. This setting works at the Analyzer level.

Trigger Source

Function

Internal

[default]

The Analyzer generates a trigger signal automatically when needed.

External

A trigger signal is a logic signal at the external trigger input (See External Trigger Settings).

Bus

The trigger signal is generated by a command from the program controlling the Analyzer via SCPI.

Manual

The trigger signal is generated by pressing the Trigger softkey in the Analyzer software.

trigger source

To set the trigger source, use the following softkeys:

Trigger > Source

Then select the required trigger source:

Internal

External

Manual

Bus

 

 

Rev.:  25.00