Please enable JavaScript to view this site.

 Vector Network Analyzers Using CMT VNA Software

The memory trace can be used for mathematical operations with the data trace. The mathematical operations are performed on complex values before they are formatted for display. The result of math operation replaces the data trace. The following mathematical operations can be performed:

Data/ Memory

Divides the measured data by the memory data.

The trace status field indicates: D/M.

Data* Memory

Multiplies the measured data by the memory data.

The trace status field indicates: D*M.

Data– Memory

Subtracts a memory data from the measured data.

The trace status field indicates: D–M.

Data+ Memory

Adds the measured data to the memory data.

The trace status field indicates: D+M.

Normalization

Pressing the Normalization softkey performs 3 steps in sequence:

1.        Saves the current data into memory.

2.        Turns on the math operation Data/ Memory (normalizes the measured data).

3.        Turns on "data only" display type.

The trace status field indicates: D/M Dat.

 

To access math operations:

Memory > Operations > Math [ Data + Memory | Data * Memory | Data / Memory | Data - Memory | None]

 

 


note

If the memory buffer function is turned on, check if the active memory trace is the desired trace for math operation (See Memory Buffer).


 